TMS instrument
The TMS instrument consists of a cryogenically cooled front-end covering the whole band of 10–20GHz through each of two pseudo-correlators, which are stabilized with a 4K reference load. In this scheme, two wide-band corrugated feedhorns, one looking at the cold blackbody calibrator and the other out of the cryostat through a microwave-transparent window, capture the reference signal and the sky emission, respectively. Both sky and load signals are fed to either correlator arm after isolating two orthogonal polarisations with a broadband Orthogonal Mode Transducer (OMT). The polar signals of both sources are correlated and decorrelated through broadband hybrid couplers before being fed to the spectrometer block. Within the correlator chain, the signals undergo several stages of amplification, as well as a filter and phase switch that ensures the needed robustness against 1/f gain variations.
TMS uses a data acquisition system (DAS) based on a System-on-Chip Field Programmable Gate Array (Xilinx ZCU208 Ultrascale) to provide the required capability to digitize and decompose the spectral band between 10–20 GHz, while filtering and correcting radio-interference signals.