EMIR in a nutshell
EMIR offers three main observing modes: Imaging, Long slit and Multi-slit spectroscopy. The spectral format is fixed for both spectroscopic observing modes using the CSU bars configuration.
Spectral Range | 0.9 - 2.5 µm [1.1 - 2.5 µm] | MOS mode | |
Top priority | MOS in K band | FOV | 6.7 x 4 arcmin (55 slitlets) |
Spectral Resol. | 5000, 4250, 4000 (JHK) | Sensitivity | K ~ 20.1 in 2h @ S/N = 5 (continuum) |
Spectral coverage | 1 single window/exp. | 1.4 x 10-18 erg/s/cm2/Å @ S/N=6 (line) | |
Detector | HAWAI2 20482 | Image Mode | |
Plate Scale | 0.2 arcsec/px | FOV | 6.7 x 6.7 arcmin |
Image quality | Θ80 < 0.3 arcsec | Sensivity | K ~ 22.8 in 1h @ S/N = 5 in 0.6 arcsec aperture |
Characteristics
|
Value
|
Pixel format
|
2048x2048
|
Pixel size
|
18 µm/pixel
|
Filling factor
|
90
|
Spectral Response
|
0.9 - 2.5 microns
|
Dark Current (77 K)
|
< 0.1 e-/s
|
Read noise (77 K)
|
< 5 e-
|
Gain
|
|
Well depth
|
|
Quantum efficiency (77K, 2.3 µm) |
75%
|
EMIR band filters cover a spectral range of 1.6 microns, from 0.9 to 2.5, with three set of filters, broad and narrow band, plus some order sorting elements. The following table provides the measured parameters at the IAC optical laboratory at room temperature at the centre of the filter and with normal incidence (Broad Band: Bb; Narrow Band: Nb and Order Sorting: Os). The transmission was measure using a collimated beam and a temperature 20 ° C. Allow for a shift of 0.7nm towards the blue to convert to cryogenic temperature.
FILTER |
Central λ |
FWHM |
Transmitance Plot |
Transmitance Data |
Type |
Y |
1031,5 |
133 |
Bb |
||
J |
1253,25 |
159,5 |
Bb |
||
H |
1629 |
281 |
Bb |
||
K-short |
2160 |
312 |
Bb |
||
K |
2234 |
322 |
Bb |
||
K spect |
2214,25 |
426,5 |
Bb |
||
Fell |
1646,75 |
29,5 |
Nb |
||
Cont. Fell |
1714,35 |
26,7 |
Nb |
||
Brackett-gamma |
2175,5 |
35 |
Nb |
||
Cont. Br-gamma |
2126,5 |
30 |
Nb |
||
H2 (1-0) |
2125,25 |
30.5 |
Nb |
||
H2(2-1) |
2249 |
29 |
Nb |
||
YJ |
1114,75 |
431,5 |
Os |
||
HK |
1929,25 |
951,5 |
Os |